Thermal Conductivity Measurement

Instrument Type:

LFA 467 Hyper Flash

Analytical Program:

Thermal conductivity measurement first temperature point -- 100 CHF/temperature point

Thermal conductivity measurement temperature point > 1 -- 40 CHF/additional temperature point

Contact Us

Introduction

LFA 467 Hyper Flash is a thermal analysis instrument that measures thermal properties of materials such as thermal conductivity, thermal diffusivity, specific heat capacity, etc. It uses the flash method to test the heat conduction properties of materials in a very short time.

Representive Results

1. Measurements of copper samples of different thicknesses are measured using a thermal conductivity tester. The tester can successfully test samples with very high thermal diffusivity, even if the sample is very thin.
2. The figure shows a comparative test of thermal diffusion coefficients of silver samples of different thicknesses. At the test temperature of 300K, the deviation of test results of samples of different thicknesses (from thin to thick) is within ±3% compared with the literature value.

Sample Requirement

1. Sample dimensions

Diameter 12.5-12.8mm, thickness 2mm, flat top and bottom surfaces. Please provide us sample density.

2. The equipment has a maximum temperature of 1200 and operates under an°Cargon gas atmosphere.

Related Products

Ar lon Milling / Polishing

Reference price: from 200 CHF/sample

Read More
SEM Morphology Characterization

Reference price: from 80 CHF/sample

Read More
Surface Area and Porosity Analyzer BET

Reference price: from 20 CHF/sample

Read More
Ultraviolet/Visible/Near Infrared Spectroscopy

Reference price: from 50 CHF/sample

Read More